This Standard Reference Material (SRM®) is primarily intended for use in the calibration and validation of procedures and methods employed in clinical analysis for the determination of electrolytes in either diluted or undiluted human serum or plasma. This SRM® can be used for calibrating direct-reading ion-selective electrode (ISE) analyzers and for quality assurance in validating secondary reference materials. A unit of NIST-956d consists of six sealed borosilicate glass ampoules of frozen human serum, two ampoules each of three different concentration levels. Each ampoule contains approximately 2 mL of human serum. Reference values for ionized calcium, phosphate and phosphorus. /// Sample value(s) - please ask for current certificate.
Inhalte
Analyt
CAS-Nummer
Konzentration
Einheit
Zusätzliche Informationen
Methode
Calcium (Ca)
[7440-70-2]
3,407 ± 0,030
mmol/L
Level 1, Coverage Factor, k = 2.00
Calcium (Ca)
[7440-70-2]
13,65 ± 0,12
mg/dL
Level 1, Coverage Factor, k = 2.00
Chloride (Cl)
94,53 ± 0,21
mmol/L
Level 1, Coverage Factor, k = 2.00
Chloride (Cl)
335,1 ± 0,8
mg/dL
Level 1, Coverage Factor, k = 2.00
Lithium (Li)
[7439-93-2]
1,965 ± 0,029
mmol/L
Level 1, Coverage Factor, k = 2.00
Lithium (Li)
[7439-93-2]
1,364 ± 0,020
mg/dL
Level 1, Coverage Factor, k = 2.00
Magnesium (Mg)
[7439-95-4]
1,471 ± 0,010
mmol/L
Level 1, Coverage Factor, k = 2.00
Magnesium (Mg)
[7439-95-4]
3,576 ± 0,025
mg/dL
Level 1, Coverage Factor, k = 2.00
Potassium (K)
[7440-09-7]
5,752 ± 0,049
mmol/L
Level 1, Coverage Factor, k = 2.00
Potassium (K)
[7440-09-7]
22,49 ± 0,19
mmol/L
Level 1, Coverage Factor, k = 2.00
Sodium (Na)
[7440-23-5]
120,0 ± 0,7
mg/dL
Level 1, Coverage Factor, k = 2.00
Sodium (Na)
[7440-23-5]
275,9 ± 1,6
mmol/L
Level 1, Coverage Factor, k = 2.00
Calcium (Ca)
[7440-70-2]
2,857 ± 0,025
mg/dL
Level 2, Coverage Factor, k = 2.00
Calcium (Ca)
[7440-70-2]
11,45 ± 0,10
mmol/L
Level 2, Coverage Factor, k = 2.00
Chloride (Cl)
108,5 ± 0,2
mg/dL
Level 2, Coverage Factor, k = 2.00
Chloride (Cl)
384,7 ± 0,8
mmol/L
Level 2, Coverage Factor, k = 2.00
Lithium (Li)
[7439-93-2]
1,207 ± 0,020
mg/dL
Level 2, Coverage Factor, k = 2.00
Lithium (Li)
[7439-93-2]
0,838 ± 0,014
mmol/L
Level 2, Coverage Factor, k = 2.00
Magnesium (Mg)
[7439-95-4]
0,961 ± 0,007
mmol/L
Level 2, Coverage Factor, k = 2.00
Magnesium (Mg)
[7439-95-4]
2,336 ± 0,016
mg/dL
Level 2, Coverage Factor, k = 2.00
Potassium (K)
[7440-09-7]
3,730 ± 0,035
mmol/L
Level 2, Coverage Factor, k = 2.00
Potassium (K)
[7440-09-7]
14,58 ± 0,14
mg/dL
Level 2, Coverage Factor, k = 2.00
Sodium (Na)
[7440-23-5]
139,3 ± 0,9
mmol/L
Level 2, Coverage Factor, k = 2.00
Sodium (Na)
[7440-23-5]
320,2 ± 2,0
mg/dL
Level 2, Coverage Factor, k = 2.00
Calcium (Ca)
[7440-70-2]
2,282 ± 0,018
mmol/L
Level 3, Coverage Factor, k = 2.00
Calcium (Ca)
[7440-70-2]
9,145 ± 0,072
mg/dL
Level 3, Coverage Factor, k = 2.01
Chloride (Cl)
122,6 ± 0,3
mmol/L
Level 3, Coverage Factor, k = 2.02
Chloride (Cl)
434,8 ± 0,9
mg/dL
Level 3, Coverage Factor, k = 2.03
Lithium (Li)
[7439-93-2]
0,455 ± 0,012
mmol/L
Level 3, Coverage Factor, k = 2.04
Lithium (Li)
[7439-93-2]
0,316 ± 0,008
mg/dL
Level 3, Coverage Factor, k = 2.05
Magnesium (Mg)
[7439-95-4]
0,434 ± 0,003
mmol/L
Level 3, Coverage Factor, k = 2.06
Magnesium (Mg)
[7439-95-4]
1,055 ± 0,008
mg/dL
Level 3, Coverage Factor, k = 2.07
Potassium (K)
[7440-09-7]
1,611 ± 0,014
mmol/L
Level 3, Coverage Factor, k = 2.08
Potassium (K)
[7440-09-7]
6,298 ± 0,054
mg/dL
Level 3, Coverage Factor, k = 2.09
Sodium (Na)
[7440-23-5]
158,7 ± 1,1
mmol/L
Level 3, Coverage Factor, k = 2.10
Sodium (Na)
[7440-23-5]
364,8 ± 2,4
mg/dL
Level 3, Coverage Factor, k = 2.11
Weiteres
Zertifikatmuster
Muster-CoA des Referenzmaterials: für die aktuelle Charge kontaktieren Sie unseren Kundendienst unter info@labmix24.com
NIST bietet über 1300 Standardreferenzmaterialien sowie Artikel und Praxisleitfäden, die die Entwicklung, Analyse und Verwendung von SRMs beschreiben.
NIST-Standardreferenzmaterialien werden von analytischen Laboratorien in Industrie, Wissenschaft und Regierung verwendet, um Handel und Gewerbe zu erleichtern und Forschung und Entwicklung voranzutreiben. Die SRMs werden vom Ingenieurlabor, dem Materialmesslabor und dem physikalischen Messlabor des NIST hergestellt. Sie stehen für chemische Zusammensetzung, physikalische Eigenschaften und technische Materialien zur Verfügung.